ORIGINAL ARTICLE
Year : 2016  |  Volume : 16  |  Issue : 2  |  Page : 118-123

Test-retest reliability of assessing cervical proprioception using cervical range of motion device


Department of Medical Rehabilitation Sciences, College of Applied Medical Sciences, King Khalid University, Abha, Kingdom of Saudi Arabia

Correspondence Address:
Dr. Ravi Shankar Reddy
Department of Medical Rehabilitation Sciences, College of Applied Medical Sciences, King Khalid University, P. O. Box No. 3665, Guraiger, Abha
Kingdom of Saudi Arabia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1319-6308.180174

Rights and Permissions

Background: Cervical proprioception plays an important role in the stability and optimal functional of the cervical spine and assessment of cervical proprioception is integral in the assessment and management of cervical spine dysfunction. Cervical range of motion (CROM) device is an effective tool, simple and cost effective tool to assess cervical proprioception. The objective of the study was to establish the test-retest reliability of a CROM in assessing cervical proprioception in asymptomatic individuals. Methods: Twenty healthy adults (mean age 37±8 years) were recruited and test retest reliability of CROM device in assessing cervical proprioception was assessed. Cervico-cephalic kinaesthetic sensibility tests - Neutral Head Positioning (NHP) and Target head Positioning (THP) tests were used to assess cervical proprioception. The test retest reliability was assessed in 2 sessions each session is separated with the other by 48 hours. Results: The test-retest reliability of measurements made with the CROM was verified with ICC values for all cervical measurements ranging between 0.66 (CI: 0.1, 0.8) for Target Head Position – rotation right to 0.93 (0.8, 0.9) (CI: 0.8-0.9) for THP – Rotation right. The correlation analysis found there are high and significant correlations between the test and retest results indicating that the reliability of the test can be established (positive correlation coefficients ranging from 0.524 – 0.863). Conclusion: CROM device can be used to quantify cervical proprioception errors with acceptable level of reliability in asymptomatic individuals.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed6129    
    Printed245    
    Emailed0    
    PDF Downloaded162    
    Comments [Add]    
    Cited by others 9    

Recommend this journal